Measuring Geometrics |
-
45 ° illumination
-
15 ° , 25 ° , 45 ° , 75 ° , 110 ° aspecular viewing
-
Angular Accuracy ±0.15 °
-
Fibre optic pick-up, coupled with DRS technology
|
Measurement Area |
12mm diameter
|
Light Source |
Gas-filled tungsten lamp, color corrected to approximately 4000 ° K
|
Illuminant Types |
C, D50, D65,D75, A, F2, F7, F11 and F12
|
Standard Observers |
2 ° and 10 °
|
Receiver |
Blue-enhanced silicon photodiodes
|
Spectral Range |
400 – 700nm
|
Spectral Interval |
28 band spectral measurement:
10nm interval from 400nm – 640nm
20nm interval from 640nm – 700nm
15nm bandwidths
|
Spectral Data Output |
Spectral reflectance values are available for output from the RS232 port for 5 angles at 10nm intervals from 400nm 700nm
|
Measurement Range |
0 to 400% reflectance
|
Measuring Time |
Approximately 2 seconds
|
Inter-Instrument Agreement |
-
0.18 D E* average on reference BCRA tile set
-
0.35 D E* maximum on any chromatic tile
-
0.15 D E* maximum on any grey tile
|
Short-Term Repeatability |
0.10 D E*ab on white ceramic (standard deviation)
|
Lamp Life |
Approximately 500,000 measurements
|
Power Supply |
Six rechargeable AA Ni-metal hydride batteries included – removable battery pack; 7.2 VDC rated @ 1400 mAh
|
AC Adapter Requirements |
-
MA68: 90 – 130VAC, 50 – 60Hz, 15W max
-
MA68X: 180 – 260VAC, 50 – 60Hz, 15W max
-
12 VDC @ 700mA
|
Charge Time |
In Instrument: 4 hours (50%), 16 hours (100%)
|
Measurements per Charge |
1,000 5-angle measurements
|
Data Storage (Five Angles) |
200 Standards, 850 Samples
|
Data Interface |
Patented bi-directional RS232, 300 – 19,200 baud
|
Display |
4-row by 20 character supertwist dot matrix LCD
|
Operating Temperature Range |
10 to 40 ° C
|