Integrated conversions |
|
|
EQUOSTAT: |
HRZ - HV - HB - HRC - HRB - HR15N - Rm |
EQUOTIP: |
or all EQUOTIP impact devices (see EQUOTIP prospectus or operating instructions). |
|
You can also enter user-specific conversions. |
Equostat unit incl. R5 probe |
|
|
Measuring Range: |
Rockwell HRZ |
equivalent hardness / tensile strength |
|
20 - 195 |
21 - 1003 HV |
|
20 - 185 |
20 - 691 HB |
|
100 - 160 |
14 - 108 HRB |
|
151 - 195 |
19 - 70 HRC |
|
140 - 195 |
58 - 94 HR15N |
|
115 - 180 |
280 - 2150 N/mm² |
Measuring Accuracy: |
±1.5 HRZ (±2.0 HRC) throughout the entire measuring range |
Resolution |
0.1 HRZ; 1 HV; 1 HB; 0.1 HRB; 0.1 HRC; 0.1 HR15N; 1 MPa (1 N/mm2) |
Total Testing Force: |
50N |
|
Penetration Depth: |
soft materials approx. 80 µm |
|
hard materials approx. 15 µm |
Conical Test Tip: |
diamond cone: point 80° |
Test Object: |
thickness: min. 0.2 mm |
|
coating thickness: min. 0.1 mm |
|
external dia.: min. 3 mm |
|
internal dia.: min. 116 mm |
|
surface finish: N6; RA = 0.8 µm |
Testing method |
static low-load Rockwell |
Available probes |
R5 (also Leeb rebound probes can be used) |
Test load |
preload 10N, total load 50N |
Min. radius of surface curvature |
1.5 mm |
|
|
|
|
Equotip 2 indicating device |
|
|
Dimensions |
175 x 180 x 80 mm |
Weight |
750 g (incl. batteries) |
Material |
shock resistant ABS plastic |
Display |
large LCD |
|
Connections: |
probe EQUOSTAT R5 / all EQUOTIP impact devices / RS232 C / AC adapter (9V DC, 0.2A) |
Data Memory: |
non-volatile memory for approx. 6000 measurements, data output to an printer and PC (on line). |
Communication |
RS232 C |
|
PC application |
Equolink 2 software included |
Battery type |
standard size "AA" cells (LR6) for about 65 h operation |
Operating temperature |
0 - 50°C |
|
Transfer of the total data memory to a PC (e.g. into an EXCEL-table). |
Date / Time: |
saved with the measurement series |
Ordering Information
350 00 105 EQUOSTAT Hardness Tester (see picture)
Includes :
-
Indicating device EQUOTIP 2,
-
probe EQUOSTAT R5 with measuring clamp (support Z0 fixed) and 1.5 m cable,
-
1 set of supports Z1/Z2/Z3/Z4
-
1 test block R3, 1 set servicing tools,
-
Carrying case 450 x 385 x 110 mm
-
Total weight 4.3 kg
350 00 106 EQUOSTAT R-H for free-hand measurements on large flat surfaces.
Includes
-
Indicating device EQUOTIP 2,
-
probe EQUOSTAT R5 with conical flat foot R-H and 1.5 m cable,
-
1 test block R3, 1 set servicing tools,
-
Carrying case 450 x 385 x 110 mm
-
Total weight 3.3 kg
Optional Accessories
EQUOLINK 2 software, download from www.proceq.com
330 00 456 Transfer cable to PC
350 74 994 Printer cable



|
|