Elcometer FD800 Bench Top Flaw Detectors

Product Description

Designed for use in the laboratory these gauges are the tool you need for all your flaw detecting needs.

The Elcometer FD800 Bench Top Flaw Detector is available in two models: FD800DL and FD800DL+.

The time corrected gain (TCG) feature automatically compensates for sound attenuation through a material, further increasing the performance of the gauge.

Within the grid batching of the FD800DL+ the user has the capability to enter ‘OBSTRUCT’ on to the grid for easy identification of inaccessible locations to measure.

The FD800DL+ has a 6Gb internal memory and an external SD slot which allows up to 64Gb with full data logging via RS232 data output to ElcoMaster® data management software.

Features of the Elcometer FD800 Bench Top Flaw Detectors

        • Blanview sunlight readable QVGA TFT colour display
        • Sizing Toolkits: DAC, AWS, TCG, DGS
        • Pulse Repetition Frequency: 8 to 333 Hz, adjustable
        • Screen Refresh Rate: 60Hz
        • Detection: Z-Cross, Flank & Peak
        • Automatic: probe zero, probe recognition, and temperature compensation
        • Measurement: Variety of modes to address a number of applications
        • Large data storage: 6Gb internal & up to 64Gb external SD slot
        • Multiple formats: Alpha numeric grid and sequential with auto identifier
        • Up to 12 hours of battery life
        • Download to ElcoMaster® data management software

       

Specifications

  • Model & Part Number FD800DL & FD800DL+
    Material thickness digits display
    B-Scan cross sectional display
    B-Scan with digits display
    Scan bar display
    Coating thickness display
    A-Scan display + Rectified, – Rectified, Full Waveform (RF)
    Flaw detection modes TRIG, DAC, AWS, TCG, Zero Crossing, Flank, Peak
    Measurement Mode1 PE, PETP (Temp Compensation),
    EE (ThruPaint™), EEV, CT (Coating) & PECT
    Measurement Rate (Thickness Mode)
    Manual: 4 readings per second
    Scan mode 32 readings per second
    Scan bar display 6 readings per second
    Measuring Range2 PE: 0.63 – 30480mm
    PETP: 0.63 – 30480mm
    EE: 1.27 – 102mm
    EEV: 1.27 – 25.4mm
    PECT: 0.01 – 2.54mm
    PECT: 0.63 – 30480mm
    Measurement Accuracy2 ± 1% or ±0.1mm whichever is the greater
    Measurement Resolution 0.01mm
    Velocity Calibration Range 256 – 16,000m/s
    Additional Features:  
    High speed scan mode
    Differential mode
    Limit alarm mode
    B-Scan display speed adjustable display speed
    Calibration setups 6 factory & 64 user-definable setups
    transferrable to and from a PC archive
    Gates 3 fully adjustable gates:
    start, stop, width & threshold
    Damping adjustable; impedance matching for
    optimising transducer performance
    Pulser type FD800DL: two adjustable square wave pulsers.
    FD800DL+: two tone burst pulsers
    Gain manual, automatic gain control (AGC)
    with 110dB range with 0.2dB resolution
    Timing precision TCXO timing with single shot
    100MHz 8bit ultra low power digitizer
    Data logging • 6Gb internal & up to 64Gb external SD slot
    • Bitmap graphic capture
    • sequential and grid logging
    • Alpha numeric batch identification
    • OBSTRUCT indicates inaccessible locations
    Calibration Options single, two point, velocity & material type
    Transducer recognition automatic
    V-path / dual path error correction automatic
    Probe zero automatic
    Flaw Detection Mode Features
    Automatic Longitudinal (straight),
    Calibration: or Shear (angle)
    Probe Types: Single Contact, Dual, Delay & Angle
    Material Velocity Table: Contains longitudinal and shear velocities for a variety of material types
    TRIG Trigonometric display of beam path,
    depth, surface distance, and curved surface correction.
    Used with angle beam transducers
    DAC Up to 8 points may be entered & used to digitally draw a DAC curve. Reference -2, -6, -10, (-6/-12), (-6/-14), (-2/-6/-10) dB. Amplitude displayed in %DAC, dB, or %FSH
    AWS Automatic defect sizing in accordance with AWS
    D1.1 structural welding code.
    AVG/DGS Automatic defect sizing using probe
    data. Stores up to 64 custom setups
    TCG Time corrected gain. 50 dB dynamic range, 20 dB
    per microsecond, up to 8 points for curve definition
    Detection Modes Zero Crossing, Flank and Peak
    Display Freeze Hold current waveform on screen
    Peak Memory Captures peak signal amplitude.
    PRF 8 to 2000Hz in selectable steps
    (8, 16, 32, 66, 125, 250, 333, 1000, 2000Hz)
    Pulse Width 40 to 400 ns. Selectable step
    options 40, 80 & 400 ns (labeled spike, thin & wide)
    Frequency Bands FD800DL & FD800DL+: Broadband 1.8 – 19 MHz (-3dB).
    Four narrow bands at 1, 2, 5, 10MHz
    FD800DL+: Additional narrow bands at 5MHz, 15MHz
    Horizontal Linearity +/- 0.4% FSW
    Vertical Linearity +/- 1% FSH
    Amplifier Linearity +/- 1 dB
    Amplitude Measurement 0 to 100% FSH, with 1% resolution
    Delay 0 – 999in (25,375mm) at steel velocity
    Display Blanview sunlight readable QVGA TFT colour display.
    115.2 x 86.4mm viewable screen
    Display Refresh Rate 60Hz
    Unit mm
    Backlight adjustable brightness
    Repeatability / Stability Indicator
    Battery Type 6 x AA alkaline
    Battery Life (approximate) 12 hours
    Low Battery Indicator
    Battery Save Mode auto
    Operating Temperature -10 to 60ºC
    Size (w x h x d) 216.0 x 165.0 x 70.0mm
    Weight (including batteries) 2.04kg
    Case Design Aluminium case design with gasket sealed end caps,
    waterproof membrane keypad
    Transducer Connector Type LEMO
    RS232 Interface Bi-directional
  • ¹ PE: Pulse-Echo Mode, EE: Echo-Echo (ThruPaint™) Mode.
    ² Measuring range & accuracy depends on material, surface conditions and the transducer selected.


    Data Sheet


    Elcometer FD800 Bench Top Flaw Detectors – Data Sheet

    Elcometer FD800 Bench Top Flaw Detectors Data Sheet

    Accessories

    Elcometer FD800 Bench Top Flaw Detectors – Accessories

    Cables & Adaptors
    TL-24030-1 T/Cable: 4′ Single Lemo 00 to BNC
    TL-24030-2 T/Cable: 4′ Single Lemo 00 to Lemo 00
    TL-24030-3 T/Cable: 4′ Single Lemo 00 to Microdot
    TL-24030-5 T/Cable: 4′ Dual Lemo 00 to BNC
    TL-24030-6 T/Cable: 4′ Dual Lemo to Lemo
    TL-24030-7 T/Cable: 4′ Dual Lemo to Microdot
    TL-24030-8 T/Cable: 4′ Dual Lemo to Microdot Single
    TL-24031 RS232 Cable (6′); DB-9 to Lemo
    TL-24032 USB to Serial Adapter
     Couplant
    TC-24034-1 Couplant: Standard; 4oz Bottle (Material Safety Data Sheet)
    TC-24034-2 Couplant: Standard; 12oz Bottle (Material Safety Data Sheet)
    TC-24034-3 Couplant: Standard; 1 Gallon (Material Safety Data Sheet)
    TC-24034-9 Couplant: Hi-Temp 510oC; 2oz Tube (Material Safety Data Sheet)
     Delay Lines
    TD-24033-1 Cone Tip Delay Line: Acrylic; 1/8″
    TD-24033-2 Cone Tip Delay Line: Acrylic; 3/16″
    TD-24033-3 Cone Tip Delay Line: Graphite; 3/16″
    TD-24033-4 Delay Tip (P): Acry; 1/16″ Dia x 0.45″ L
    TD-24033-5 Delay Tip (P): Acry; 1/8″ Dia x 0.45″ L
    TD-24033-6 Delay Tip: Acrylic; 1/4″ Dia x 1/2″ L
    TD-24033-7 Delay Tip: Acrylic; 1/4″ Dia x 3/8″ L
    TD-24033-8 Delay Tip: Graphite; 1/4″
    Other Accessories
    TZ-24035 6″ Ext Wand for S/E Microdot Transducers

     

     

    Instruction Manual

    Elcometer FD800 Bench Top Flaw Detectors – Instruction Manual