About this product

The Gamry Interface 1010E is a potentiostat, galvanostat and zero resistance ammeter (ZRA) for the electrochemistry laboratory. Gamry lists it as the general purpose instrument of the Interface 1010 range, so it is the variant to specify when the work is not narrowed to a single technique.

It carries the widest envelope of the three 1010 variants. Maximum current is ±1 A across 9 current ranges (10 nA to 1 A, or 11 ranges with internal gain applied), maximum applied potential is ±12 V, and minimum current resolution is 3.3 fA. EIS runs from 10 µHz to 2 MHz, which is the clearest separation from the Interface 1010B and Interface 1010T: both of those stop at 20 kHz.

The control amplifier gives ±20 V compliance at 1 A with five speed settings. The electrometer has an input impedance of >10^12 Ω, an input current below 20 pA, bandwidth above 15 MHz and CMRR above 86 dB at 10 kHz.

It takes 2, 3 or 4 cell connections, floats clear of earth ground, and weighs 2.4 kg.

Values listed as typical are representative rather than guaranteed.

Gamry

Gamry Interface 1010E Potentiostat / Galvanostat / ZRA

SKU · Interface 1010E

The general purpose instrument of the Gamry Interface 1010 range. Maximum current ±1 A across 9 current ranges, maximum applied potential ±12 V, and EIS from 10 µHz to 2 MHz. It takes 2, 3 or 4 cell connections, is isolated from earth, and weighs 2.4 kg.

    Indicative price
    POA
    Specialist-built quote, by close of business
    Or ask a quick question on WhatsApp
    Used by inspection teams in
    RefiningMarineMiningBridge / civil
    Specifications

    Configuration

    Potentiostat
    Yes
    Galvanostat
    Yes
    Zero Resistance Ammeter
    Yes
    Cell Connections
    2, 3, or 4
    Floating (isolated from earth ground)
    Yes

    System

    Max. Current
    ±1 A
    Current Ranges
    9 (10 nA - 1 A)
    Current Ranges (including internal gain)
    11
    Min. Current Resolution
    3.3 fA
    Max. Applied Potential
    ±12 V
    Rise Time
    <1 µs
    Min. Time Base
    10 µs
    Noise and Ripple
    <20 µV rms

    EIS measurement

    Frequency Range
    10 µHz - 2 MHz
    EIS Accuracy
    See accuracy contour plot
    AC Amplitude
    2.33 V maximum; 50 µV minimum

    Control amp

    Compliance Voltage
    ±20 V (@ 1A)
    Output Current
    >±1 A
    Speed Settings
    5
    Unity Gain Bandwidth
    980, 260, 40, 4, 0.4 kHz

    Electrometer

    Input Impedance
    >10^12 Ω
    Input Current (typical)
    <20 pA
    Bandwidth
    >15 MHz
    Common Mode Rejection Ratio (CMRR)
    >86 dB (10 kHz); >60 dB (1 MHz)

    Potential

    Applied Accuracy
    ±1 mV ±0.2% of setting
    Applied Resolution
    200 µV, 50 µV, 12.5 µV/bit
    Measured Accuracy
    ±1 mV ±0.3% of reading
    Measured Resolution
    1 µV, 10 µV, 100 µV, 400 µV/bit

    Current

    Applied Accuracy
    ±5 pA ±0.3% of range
    Applied Resolution
    0.0033% full-scale/bit
    Measured Accuracy
    ±5 pA ±0.3% of range
    Measured Resolution
    0.0033% full-scale/bit
    Measured Current Bandwidth
    >10 MHz (1A - 100 µA); >1.5 MHz (10 µA); >0.15 MHz (1 µA)

    Physical

    Dimensions
    24 (W) x 6 (H) x 27 (D)
    Weight
    2.4 kg
    Contact us

    Quote on the Gamry Interface 1010E Potentiostat / Galvanostat / ZRA by close of business.

    Tell us about your application and we will send a quote with lead time and accessories.

    Not sure this is the right model for your job? for a quick recommendation.

    Request a quote

    Request a quote

    Tell us about your application and we will send a quote with lead time and accessories.

    Contact us