The Elcometer 6910 Setaflash 'Series 8' Active-Cool Closed Cup Tester (part number K6910M021) is a closed cup flash point tester with a temperature range of -30 degrees C to 135 degrees C that requires no external cooling. Flash point determines the lowest temperature at which a liquid gives off enough vapour to ignite, which sets how a substance is classified, transported and stored, so accurate flash point testing is central to chemical safety and regulatory compliance.
The Elcometer 6910 Setaflash uses Peltier cell technology for both heating and cooling, giving precise temperature control and rapid cool down after a test. It tests a small sample of 2 ml or 4 ml depending on the method and offers Flash/No Flash and Ramp modes, along with an electric ignitor. In Rapid Equilibrium mode a test takes 1 minute below 100 degrees C and 2 minutes above, while Ramp mode typically takes 7 minutes; cool down from 80 degrees C to 20 degrees C takes 4 minutes. The aluminium cup unit of the Elcometer 6910 Setaflash measures 300 x 340 x 380 mm and weighs 8 kg, with 1 GB of internal memory and an integrated USB port for results. A calibration certificate is supplied as standard.
The Elcometer 6910 Setaflash can be used in accordance with ASTM D3278, D3828, D7236 and E502, BS 2000-523, EU CLP Regulations EC No 1272/2008, EPA 1020 A and B, ISO 3679 and 3680, and UN Class 3 Non-viscous Flammable Liquids. Supplied by BAMR, South Africa's authorised Elcometer distributor.
Elcometer 6910 Setaflash Series 8 Active Cool Closed Cup Tester
The Elcometer 6910 Setaflash 'Series 8' Active-Cool Closed Cup Tester offers the same features as the Setaflash 'Series 8' Closed Cup Tester but is designed with a temperature range of -30°C to 135°C with no external cooling required.
- Small sample size 2ml or 4ml
- Flash/No Flash & Ramp Modes
- Electric ignitor
- Active-Cool electronic Peltier cooling
- 1GB internal memory & integrated USB port
Specifications
| Part Number | Description | Certificate |
|---|---|---|
| K6910M021 | Elcometer 6910 Setaflash 'Series 8' Active-Cool Closed Cup Tester | ● |
| Temperature Range | -30°C to 135°C | |
| Sample Size | 2ml or 4ml according to method | |
| Test Mode | Rapid Equilibrium and Ramp | |
| Test Duration Rapid Equilibrium Mode | 1 minute below 100°C 2 minutes above 100°C | |
| Test Duration Ramp Mode | Typically, 7 minutes | |
| Cool Down Time | 80°C to 20°C in 4 minutes | |
| Cup Material | Aluminium | |
| Dimensions | 300 x 340 x 380mm | |
| Weight | 8kg |
Specifications
Specifications
| Part Number | Description | Certificate |
|---|---|---|
| K6910M021 | Elcometer 6910 Setaflash 'Series 8' Active-Cool Closed Cup Tester | ● |
| Temperature Range | -30°C to 135°C | |
| Sample Size | 2ml or 4ml according to method | |
| Test Mode | Rapid Equilibrium and Ramp | |
| Test Duration Rapid Equilibrium Mode | 1 minute below 100°C 2 minutes above 100°C | |
| Test Duration Ramp Mode | Typically, 7 minutes | |
| Cool Down Time | 80°C to 20°C in 4 minutes | |
| Cup Material | Aluminium | |
| Dimensions | 300 x 340 x 380mm | |
| Weight | 8kg |
Standards
Standards
Accessories
Accessories
| KT006910N009 | Certified Flash Point Material |
Frequently asked questions
Do I need external cooling equipment with the Series 8 Active-Cool?
No. The built-in Active-Cool system manages the full temperature range of -30°C to 135°C without requiring external cooling units. This simplifies your lab setup and reduces capital expenditure.
How long does a typical test take?
Test duration depends on the mode and temperature range. Rapid Equilibrium mode takes 1 minute below 100°C or 2 minutes above 100°C. Ramp mode typically takes 7 minutes. Your test method will specify which to use.
How quickly can I run consecutive tests?
The tester cools from 80°C to 20°C in 4 minutes, enabling reasonably fast turnaround between tests. This timeframe supports good throughput in busy QC environments.
What sample size do I need for each test?
Sample volume is either 2ml or 4ml, depending on your test method. Your relevant standard or internal procedure will specify the correct volume for your application.
Can this tester handle both extreme cold and high-temperature testing?
Yes. The range spans -30°C to 135°C, allowing a single instrument to test materials across cold climates, ambient conditions, and elevated-temperature applications without additional equipment.
Quote on the Elcometer 6910 Setaflash Series 8 Active Cool Closed Cup Tester by close of business.
Tell us about your application and we will send a quote with lead time and accessories.
Not sure this is the right model for your job? for a quick recommendation.
Request a quote
Request a quote
Tell us about your application and we will send a quote with lead time and accessories.
Shop online